Table 3 - Calibration Weight Table - Conitnued

“1 / 4”

- 1 / 4 Capacity

Mode (Capacity

KG)

Capacity LB

Calibration

Weight KG

Calibration

Weight LB

Acceptable

Span Range

CS_ _6 (6KG)

12

N/A

N/A

N/A

CS_15 (15KG)

30

2 KG

5 LB

16000-24000

CS_30 (30KG)

60

5 KG

10 LB

20000-30000

CS_60 (60KG)

150

10 KG

20 LB

16000-30000

CS_150

(150KG)

300

20 KG

50 LB

16000-24000

CS_300

(300KG)

600

N/A

N/A

N/A

* Grey Shaded areas indicate capacities only available for Model 3255.

SEALING YOUR SCALE FOR WEIGHTS AND MEASURES APPROVAL

If your scales requires sealing by your local weights and measures authorities, please follow the

instructions shown below to seal your scale.

*

IMPORTANT NOTE:

Once you have set the scale to the SEALED mode, access to the

configuration menu is limited to changing the

“A.OFF”

settings

only

. To enter into the

configuration menu for the other features or to calibrate your scale, you will be required to follow

the steps shown on page X, “

Calibration Procedure Sealed”.

FUNCTION

SCALE RESPONSE

1. Follow step 1-13 of the

“Calibration Procedure Non-Sealed”.

2. Press and hold the

SET

key for three

seconds to enter into the configuration mode.

Scale displays “Mode”

3. Press the

CALL

key 7 times to “

trad”,

press the

Set

key.

Scale displays “OFF”.

4. Press the

CALL

key once

Scale displays “ON”.

5. Press the

SET

key to active the sealing mode

Scale displays “trAd”

6. Press the

CALL

key once

Scale displays “_END_”.

7. Press the

SET

key to exit and return to normal weighing mode.

Once you have set the scale to the SEALED mode, access to the configuration menu is limited to

changing the

“A.OFF”

settings

only

. To enter into the configuration menu for the other features or

to calibrate your scale, you will be required to follow the steps shown on page X, “

Calibration

Procedure Sealed”.