Table 3 - calibration weight table - conitnued – Salter Brecknell C3255 User Manual
Page 14

1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
Page 14 of 18
Table 3 - Calibration Weight Table - Conitnued
“1 / 4”
- 1 / 4 Capacity
Mode (Capacity
KG)
Capacity LB
Calibration
Weight KG
Calibration
Weight LB
Acceptable
Span Range
CS_ _6 (6KG)
12
N/A
N/A
N/A
CS_15 (15KG)
30
2 KG
5 LB
16000-24000
CS_30 (30KG)
60
5 KG
10 LB
20000-30000
CS_60 (60KG)
150
10 KG
20 LB
16000-30000
CS_150
(150KG)
300
20 KG
50 LB
16000-24000
CS_300
(300KG)
600
N/A
N/A
N/A
* Grey Shaded areas indicate capacities only available for Model 3255.
SEALING YOUR SCALE FOR WEIGHTS AND MEASURES APPROVAL
If your scales requires sealing by your local weights and measures authorities, please follow the
instructions shown below to seal your scale.
*
IMPORTANT NOTE:
Once you have set the scale to the SEALED mode, access to the
configuration menu is limited to changing the
“A.OFF”
settings
only
. To enter into the
configuration menu for the other features or to calibrate your scale, you will be required to follow
the steps shown on page X, “
Calibration Procedure Sealed”.
FUNCTION
SCALE RESPONSE
1. Follow step 1-13 of the
“Calibration Procedure Non-Sealed”.
2. Press and hold the
SET
key for three
seconds to enter into the configuration mode.
Scale displays “Mode”
3. Press the
CALL
key 7 times to “
trad”,
press the
Set
key.
Scale displays “OFF”.
4. Press the
CALL
key once
Scale displays “ON”.
5. Press the
SET
key to active the sealing mode
Scale displays “trAd”
6. Press the
CALL
key once
Scale displays “_END_”.
7. Press the
SET
key to exit and return to normal weighing mode.
Once you have set the scale to the SEALED mode, access to the configuration menu is limited to
changing the
“A.OFF”
settings
only
. To enter into the configuration menu for the other features or
to calibrate your scale, you will be required to follow the steps shown on page X, “
Calibration
Procedure Sealed”.