CALIBRATION PROCEDURE – NON- SEALED- Conitnued
FUNCTION
SCALE RESPONSE
11. Load the appropriate calibration weight.
“XXXXXX”
See Table 3 for valid calibration
weights used, verify weight placed
on platform is with acceptable span
range.
12. Press the
TARE
key to begin calibration
“
-CAL-
“
13. The scale will exit out of calibration mode
and return to normal weighing mode.
“
Scale will display weight used for
calibration
“
If calibration process does not correct any weighing issues, contact the service department for
technical support at 1-800-458-7062.
Table 3 - Calibration Weight Table
“FULL” – Full capacity
Mode (Capacity
KG)
Capacity LB
Calibration
Weight KG
Calibration
Weight LB
Acceptable
Span Range
CS_ _6 (6KG)
12
N/A
N/A
N/A
CS_15 (15KG)
30
10 KG
20 LB
80000-120000
CS_30 (30KG)
60
20 KG
40 LB
80000-120000
CS_60 (60KG)
150
50 KG
100 LB
100000-150000
CS_150
(150KG)
300
100 KG
200 LB
80000-120000
CS_300
(300KG)
600
N/A
N/A
N/A
* Grey Shaded areas indicate capacities only available for Model 3255.
“1 / 2”
– 1 / 2 Capacity
Mode (Capacity
KG)
Capacity LB
Calibration
Weight KG
Calibration
Weight LB
Acceptable
Span Range
CS_ _6 (6KG)
12
N/A
N/A
N/A
CS_15 (15KG)
30
5 KG
10 LB
40000-60000
CS_30 (30KG)
60
10 KG
20 LB
40000-60000
CS_60 (60KG)
150
20 KG
50 LB
320000-60000
CS_150
(150KG)
300
50 KG
100 LB
40000-60000
CS_300
(300KG)
600
N/A
N/A
N/A