Taylor-Wharton CS Series User Manual
Page 20
18
CS SERIES
View Only
The View Only menu selection allows the user to view settings but will not allow
changes to any settings. This restriction is designated with a lock symbol. Refer to
Figure 8.1 Menu System - View Only on page 22 to navigate the “View Only” menu.
Change Settings
The Change Settings menu selection allows the user to change operational settings
for the control system. Refer to Figure 8.2 Menu System - Change Settings on page
23 to navigate through any changes you need to make.
Below are the Menu Settings that are available in both the “View
Only” and “Change Settings” modes.
• TEMPERATURE
– Thermocouple Select: Allows the user to turn on/off thermocouples for
temperature measurement.
– Calibrate
• Calibrate Thermocouple 1: Allows the user to calibrate the control for use
with thermocouple #1.
• Calibrate Thermocouple 2: Allows the user to calibrate the control for use
with thermocouple #2.
• Restore Calibration: Resets the calibration to the data stored during factory
calibration.
• Set Ambient: Ambient temperature is set at the factory and should not be
changed in the field.
– Test Temperature System
• Check Thermocouples: Gives status of temperature sensors.
• Test Temp Alarms: Allows the user to manually test temperature alarms for
the thermocouples.
– Control by Temperature
• Temp Control On/Off: Enables/Disables temperature control.
• Temp Control Range: Allows the user to set the range for temperature control.
– Temperature Alarms
• High Temperature Alarm: Allows the user to set the high temperature
alarm for both thermocouples. Settings [00° C to -1900° C]
• Low Temperature Alarm: Allows the user to set the low temperature alarm
for both thermocouples. Settings [00° C to -1900° C]
• Time with no Temp alarm: Displays the time since the last high or low
temperature alarm.
– Temperature Units: Allows the user to choose the units of measure for
temperature.
• LEVEL
– Thermistor Status: Gives the user the status for the 8 thermistors on the level
sensor assembly and the freezeguard sensor installed at the valve.
– Sensor Positions: Allows the user to assign the thermistors for low level alarm,
start fill, stop fill, high level alarm and the sensor offset.
– Valve Open Duration: Allows the user to set the fixed times for valve open
and valve closed durations.