Humboldt GeoGauge

®

Verifier Mass Guide

For use with C series Humboldt GeoGauge

Purpose

To verify or check the operation of the C series Humboldt H-4140 GeoGauge by

running a measurement on top of the Verifier Mass. It is not intended to

calibrate the GeoGauge.
Equipment Required

H-4140.20 Verifier Mass Assembly (10 kg with installed rubber isolation mounts)
Operation
1) Place the Verifier Mass Assembly on a relative flat and rigid surface. The

surface at this location should be approximately level. A concrete floor that

is generally crack-free and well supported is ideal.

2) Wipe clean the ring foot at the bottom of the GeoGauge. Clean the

v-groove around the seal between the foot and gauge body. Inspect the

ring foot surface for deep gouges, nicks, protrusions or raised nicks. The

ring foot should be relatively flat. Remove excessive protrusions or raised

nicks by setting the GeoGauge on a flat abrasion sheet placed on a flat steel

plate. Rotate the GeoGauge via its handle 2 – 3 times until the protrusions

are approximately flat with the ring foot surface. Wipe off the abrasion dust

from the ring foot.

3) Turn on GeoGauge. Set the GeoGauge to display stiffness in

S.I. (metric) units.

4) Smear a small amount of any lubricating oil on the shoulder of

the Verifier Mass.

5) Gently set the GeoGauge’s ring foot in position over the shoulder of

the Verifier Mass.

6) Rotate the gauge on the mass a random amount.
7) Firmly press the MEAS button to make a measurement of the

Verifier Mass stiffness.

8) After 75 seconds the measured stiffness will be displayed. Record

the stiffness.

9) Remove the GeoGauge from the Verifier Mass. Reset it back onto the

Verifier Mass. When repeating measurements, it is important to remove

the GeoGauge from the Verifier Mass between measurements to account for

placement and operator bias. Place the GeoGauge on the Verifier at different

rotational orientation each time.