7.3 Command Language for the Model 550 Microplate Reader

SYNTAX:[ ]

device name: “EIA. READER” -Model 550 Microplate READER

1. : “ID”-request device ID. 550 READER id; “550”; no arguments

:

“ERE”

:

“0550”: Model 550 Microplate READER ID code]

2. : “AQ” - Acquire remote control of the device and lock front panel

key pad; no arguments
When acquired, the reader will be in the Remote Control mode. The
LCD will display Remote Control mode, and the printer will automat-
ically be deactivated. The reader will stay in the remote control mode
until the release (RL) command is received or the STOP button on the
front panel is pressed.

: “ERE”

3. : “RL”- Release remote control of the device and release front panel key

pad; no arguments

: “ERE”

4. : “RS”- Reset device to power up configuration local mode; no arguments

: “ERE”

5. : “MR”- Transmit maintenance report; no arguments

: “ERE”
< >: maintenance records;

“On/off:” <#on/off> :Number of times the reader was turned on.
“Hours:” <#hours> :Number of hours the reader was on.
“Plates:” <#plates> :Number of plates read.

Note: <#on/off>,<#hours> and <#plates> are 4 ASCII encoded decimal values. This
information is kept in battery backup RAM.

6. : “RM”- Reset maintenance report values; no arguments

: “ERE”

7. : “RWELL”- Read a plate well; arguments , ,

:

cl - Column number ASCII encoded decimal (1-12)

rw - Row number ASCII encoded decimal (1-8)
wp1 - Measurement filter position ASCII encoded decimal (1-4)
wp2 - Reference filter position ASCII encoded decimal (1-4)
Optional; only present for Dual wavelength reading

: “ERE”

: Absorbance values at measurement wavelength (ASCII encoded

floating point format) ex.” 1.234”

:

Absorbance values at reference wavelength (ASCII encoded floating
point format) ex.”1.234”

33