Bio-Rad Model 550, Ultramark, and Benchmark Microplate Reader Accessories User Manual
Page 36

7.3 Command Language for the Model 550 Microplate Reader
SYNTAX:
device name: “EIA. READER” -Model 550 Microplate READER
1.
“ERE”
“0550”: Model 550 Microplate READER ID code]
2.
key pad; no arguments
When acquired, the reader will be in the Remote Control mode. The
LCD will display Remote Control mode, and the printer will automat-
ically be deactivated. The reader will stay in the remote control mode
until the release (RL) command is received or the STOP button on the
front panel is pressed.
3.
pad; no arguments
4.
5.
< >: maintenance records;
“On/off:” <#on/off>
“Hours:” <#hours>
“Plates:” <#plates>
Note: <#on/off>,<#hours> and <#plates> are 4 ASCII encoded decimal values. This
information is kept in battery backup RAM.
6.
7.
cl - Column number ASCII encoded decimal (1-12)
rw - Row number ASCII encoded decimal (1-8)
wp1 - Measurement filter position ASCII encoded decimal (1-4)
wp2 - Reference filter position ASCII encoded decimal (1-4)
Optional; only present for Dual wavelength reading
floating point format) ex.” 1.234” Absorbance values at reference wavelength (ASCII encoded floating 33
point format) ex.”1.234”