q. Press NEXT to move the cursor to the second digit. Use either ARROW to set
the second digit.
r. Repeat step q until all four digits are set correctly.
s. Press NEXT to return the cursor to the # position. Use either ARROW to change
the ‘1’ to ‘2’.
t. Continue using NEXT and ARROW to set the concentrations of each standard.
When all the standards have been defined, press ENTER twice and the display
will revert to
Note: If during the setting of concentrations of the standards, the error message
appears, it indicates that the standard concentrations are not being input in
an ascending or descending manner.
5.
Editing Sample Parameters
a. Use either ARROW to move the cursor under the check box next to E Sample.
b. Press SELECT to darken the check box.
c. Press ENTER. The display will change to
d. To set the number of samples, use either ARROW to move the cursor next to
SAMPLEs.
e. Press SELECT to darken the check box.
f. Press ENTER and the screen will change to
where xy represents the currently defined number of samples.
g. The cursor will be under the x position. Use either ARROW to set the x digit.
h. Press NEXT to move the cursor to the y position. Use either ARROW to set the
y digit.
i. Press ENTER to seal the choices. The display will revert to
25
Err: Conc. Values
Press any key
■
■
E STD
■
■
E Sample
■
■
Print Format
■
■
SAMPLEs
■
■
Sample Repl
■
■
SAMPLEs
■
■
Sample Repl
Set Sample:
SAMPs = xy (0~88)