Set Report Types Mode
Choose the reports to be generated and automatically printed from within this mode.
1. Press MODE until the display reads
2. Press either ARROW until the display reads
3. Press ENTER. The display changes to read
4. To select a report type, use either ARROW to move the cursor under the check
box associated with the report type, then press SELECT. The check box will darken.
To deselect a choice, move the cursor under the darkened check box and then press
SELECT.
5. When the check boxes next to each desired report type have all been darkened, press
ENTER to seal the choices. The display will return to the previous screen.
Set Blanks Mode
In this series of screens, the positions of blank wells are defined or edited. Blanks are
required for all reports other than Raw. Once blank positions are defined, they are retained in
memory until they are edited through the Set Blanks Mode or until battery failure.
1. Press MODE until the display reads
2. Press either ARROW until the display reads
16
>PRINT REPORT
MODE
>SET REPORT
TYPES MODE
>SET REPORT
TYPES MODE
■
■
Raw
■
■
Abs
■
■
Limit
■
■
Matr
■
■
Cut
■
■
Conc
>PRINT REPORT
MODE
>SET BLANKS
MODE