j. To set the number of replicates for each sample, use either ARROW to move the
cursor under the check box next to Sample Repl.
k. Press SELECT to darken the check box.
l. Press ENTER and the screen will change to
where y is the currently set number of sample replicates.
m. Use either ARROW to set y to the desired number of replicates and then press
ENTER twice to return the screen to
6. Printing the Concentration format. The concentration format printout will indicate
the locations of the blanks, the standards, and the samples. It also prints the concentrations
of the standards.
a. Use either ARROW to move the cursor under the check box next to Print Format.
b. Press SELECT to darken the check box.
c. Press ENTER. Printing will begin immediately.
d. When printing is completed, the screen will return to.
Printing Reports
All reports selected in the SET REPORT TYPES mode will be printed automatically at
the end of plate reading. Since the memory buffer stores plate data until another plate is read,
other reports may be generated after a plate is read. Once another plate is read, all previous-
ly stored data are lost. If a plate reading is aborted before completion, the stored data are
saved and none of the new data are saved. It is also possible to print a report showing system
configuration. The data include: the location of blanks, reading mode, limit settings, cutoff set-
tings, concentration settings, and sample positions.
1. To print additional data reports
a. Enter the SET REPORT TYPES mode and select the additional reports desired
(See above).
b. Press MODE until the display shows
26
Set STD Repl:
REPL = y (1~2)
■
■
E STD
■
■
E Sample
■
■
Print Format
>SET CONC.
MODE
>PRINT REPORT
MODE