5. Filter selection:
a. Use either ARROW to move the cursor under the check box next to Filt.
b. Press SELECT to darken the check box next to Filt.
c. Press ENTER. The display changes to read
where x and y refer to filters 1, 2, 3, or 4.
d. The cursor will be under the Measurement filter. Use either ARROW to select
the filter to be used for measurement.
e. Press NEXT to move the cursor to the Reference filter position. Use either
ARROW to select the filter to be used for reference. Note: a Reference filter need
only be chosen for dual wavelength measurements.
f. Press ENTER to finalize the selection and the display will return to the previous
screen.
6. Mixing:
a. Use either ARROW to move the cursor under the check box next to Mix.
b. Press SELECT to darken the check box next to Mix.
c. Press ENTER. The display changes to read
d. Use either ARROW to move the cursor under the desired check box.
f. Press ENTER to finalize the selection and the display will return to the previous
screen
7. Press ENTER again to go back to the original screen.
15
Set Filter:
Mes.=#x Ref.=#y
Set Analysis
■
■
D/S,
■
■
Filt and
■
■
Mix
Set Mixing:
■
■
Yes/
■
■
No
Set Analysis
■
■
D/S,
■
■
Filt and
■
■
Mix
>SET ANALYSIS
MODE