e. When the desired row is chosen, press NEXT to move the cursor to the column

position.

f. Again, use either ARROW to step up or down between 1 and 12 until the desired

column number is displayed.

g. The check box on the bottom line of the display will be darkened if the chosen well

is presently defined as blank. If the well is not presently defined as blank, then
the check box is empty. To change the status of a well, press SELECT. If the
check box had been darkened, it will be emptied, and if the check box started out
as empty, it will be darkened.

h. Press ENTER to seal the selection/editing of the blank well(s). The display will

return to the previous screen.

8. Press ENTER to return the screen to

Set Limits Mode

If either the Matrix or Limit report is selected, then Upper and Lower limits must be

defined in the Set Limits mode. The Upper limit may not exceed 3.500, nor may the Upper
limit be less than the Lower limit. If either of these situations occurs, an error message will be
displayed.

1. Press MODE until the display reads

2. Press either ARROW until the display reads

3. Press ENTER. The display changes to read

19

Clear Blanks

Col

Row

Well

>SET BLANKS

MODE

>PRINT REPORT

MODE

>SET LIMITS

MODE

Err: Max=3.500

Press any key

Err: Lower>=Upper

Press any key

Upper= 3.500

Lower= 0.000