beautypg.com

Prepare reagents 12, Prepare reagents, Setting up the bio-plex handheld magnetic washer – Bio-Rad Human Metabolic and Hormone Assays User Manual

Page 14: Setting up a vacuum manifold

background image

12

Setting Up the Bio-Plex Handheld Magnetic Washer

Place an empty flat bottom plate on the magnetic washer by sliding
it under the retaining clips. Push the clips inward to secure the plate.
Make sure the plate is held securely. If needed, the clips can be adjusted
for height and tension. For detailed instructions, refer to the user guide
(bulletin #10023087).

Setting Up a Vacuum Manifold

Calibrate the vacuum manifold by placing a standard 96-well flat bottom
plate on the unit and adjusting the pressure to –1 to –3

"

Hg. In general,

100 µl liquid should take 3–4 sec to clear the well. For more detailed
instructions, refer to bulletin #10005042.

4. Prepare Reagents

1. Reconstitute the following lyophilized reagents in dH

2

0 before use

according to the table below.

a. Allow vial to sit at room temperature for a minimum of 5 min, not to

exceed 30 min.

b. Mix by vortexing at a medium setting.

2. Bring the 10x assay buffer to room temperature (RT).

a. Mix by inversion to ensure all salts are in solution.

b. Prepare 1x assay buffer — dilute 1 part 10x assay buffer (60 ml) with

9 parts of dH

2

0 (540 ml).

Reagent

dH

2

O Volume

Standards mix

150 µl

Control 1

100 µl

Control 2

100 µl

Blocking buffer

1.5 ml

Standard diluent

1.0 ml

Detection antibodies

4.8 ml

Table 4. Reagent volume.