Bio-Rad Bio-Plex Pro™ Rat Cytokine, Chemokine, and Growth Factor Assays User Manual
Page 5
3
3
Biomarker
of Interest
Magnetic Bead
Capture
Antibody
Biotinylated
Detection
Antibody
Phycoerythrin
Fluorescent
Reporter
Streptavidin
Fig. 1. Bio-Plex sandwich immunoassay.
Data Acquisition and Analysis
Data from the reactions are acquired using a Bio-Plex system or similar
Luminex-based reader. When a multiplex assay suspension is drawn into
the Bio-Plex 200 reader, for example, a red (635 nm) laser illuminates the
fluorescent dyes within each bead to provide bead classification and thus
assay identification. At the same time, a green (532 nm) laser excites PE
to generate a reporter signal, which is detected by a photomultiplier tube
(PMT). A high-speed digital processor manages data output, and
Bio-Plex Manager
™
software presents data as median fluorescence
intensity (MFI) as well as concentration (pg/ml). The concentration of
analyte bound to each bead is proportional to the MFI of reporter signal.
Using Bio-Plex Data Pro
™
software, data from multiple instrument runs
can be combined into a single project for easy data management, quick
visualization of results, and simple statistical analysis.