Bio-Rad Bio-Plex Pro™ Rat Cytokine, Chemokine, and Growth Factor Assays User Manual
Page 46
44
2. Determine the volume of 1x diabetes and cytokine detection antibodies needed.
a) Each well requires 25 μl of detection antibodies (1x): _______ x 25 μl = ______ μl
1 13
b) Include 25% excess to ensure enough volume: _______ μl x 0.25 = ______ μl
13 14
c) Total volume of 1x detection antibodies: ______ μl + ______ μl = ______ μl
13 14 15
d) Volume of 20x detection antibodies required from each diabetes tube(s):
______ μl / 20 = ______ μl
15 16
e) Volume of 10x detection antibodies required from each cytokine tube(s):
______ μl / 10 = ______ μl
15 17
f) Total volume of diabetes detection antibodies stock required: ____ μl x ____ =_____ μl
16
2 18
g) Total volume of cytokine detection antibodies stock required: ____ μl x ____ =_____ μl
17 3 19
h) Total volume of combined detection antibodies required:
_____ μl + ______ = _____ μl
18
19 20
i) Volume of detection antibody diluent required: ______ μl – _____ μl = _____ μl
15 20 21
3. Determine the volume of 1x streptavidin-PE needed.
a) Each well requires 50 μl of streptavidin-PE (1x): _________ x 50 μl = _________ μl
1 15
b) Include 25% excess to ensure enough volume: _______ μl x 0.25 = ______ μl
15 16
c) Total volume of 1x streptavidin-PE: ______ μl + ______ μl = ______ μl
15 16 17
d) Volume of 100x streptavidin-PE required: _______ μl / 100 = ______ μl
17 18
e) Volume of assay buffer required: ______ μl – ______ μl = __________ μl
17 18
19