beautypg.com

Bio-Rad Components for Older Model Spot Cutter User Manual

Page 7

background image

2.2

Remove the Door

The door is easily removed at the hinges for easier access to the interior while installing the lights,
and for placing it over the spot cutter during installation.

1. Hold the door, or have someone else hold the door, while disconnecting the hinges.

2. To disconnect the hinges, push the two catches vertically in towards the center of the

hinge, and then turn the metal tab/lever away from the hinge into the locking slot.

Fig. 2. Opening the hinge.

Fig. 3. Disconnected door.

2.3

Installing the UV Lights

1. Remove one of the two UV light assemblies from the box. Note that one is made for the

left side, and one for the right, of the enclosure. These are identified by their orientation
with the electrical connection cable. Hold the light assembly with the bracket toward the
wall of the Enclosure. The electrical cable should be at the back end of light assembly,
toward the back of the Enclosure.

Fig. 4. Right side light assembly.

2. Set the right light assembly so that the back light assembly bracket slots into the back

mounting bracket on the right inside of the Enclosure.

4

Connection cable

Back bracket

Front bracket

UV light filter