Bio-Rad PROTEAN Plus Dodeca Cell User Manual
Page 4
2
Specifications of the Buffer Recirculation Pump
Number of channels
1
Flow rate range
0 - 1.5 liters/min, as measured with 3/8" tubing and
water. Flow rate will vary with tubing diameter,
buffer viscosity, and line voltage.
Maximum counterpressure
~ 10 psi
Motor
Self-priming oscillating pump
Power requirements
120 VAC, 60 Hz
220 VAC, 50 Hz
Maximum ambient temperature limit 40
°
C
Fuses required
100/120 V use .25 A type T
220/240 V use .125 A type T
Dimensions
19 cm x 12.7 cm x 18 cm (W x D x H)
Weight
2.9 kg
Note: Dodeca cell components are not compatible with acetone or ethanol. Use of organic
solvents voids all warranties.
1.2 Safety
Power to the PROTEAN Plus Dodeca cell is supplied by an external DC voltage power
supply (not included). The output of the power supply must be isolated from external ground
to insure that the DC voltage output floats with respect to ground. All Bio-Rad power supplies
meet this important safety requirement. Regardless of the power supply used, the maximum
specified operating parameters for the PROTEAN Plus Dodeca cell are as follows:
1000 VDC
voltage limit
250 watts
power limit
40 °C
ambient temperature limit
The current to the cell enters the unit through the lid assembly that provides a safety interlock
to the user. Always turn off the power supply before removing the lid. Do not attempt to use the
cell without the safety lid.
Important: This Bio-Rad product is designed and certified to meet *EN61010-1 safety standards.
Certified products are safe to use when operated in accordance with the instruction manual. This
instrument should not be modified or altered in any way. Alteration of this instrument will
• Void the warranty
• Void the EN61010-1 certification, and
• Create a potential safety hazard
Bio-Rad is not responsible for any injury or damage caused by use of this instrument for
purposes other than those for which it is intended or by modifications of the instrument not
performed by Bio-Rad or an authorized agent.
*EN61010-1 is an internationally accepted electrical safety standard for laboratory instruments.