beautypg.com

Bio-Rad PROTEAN Plus Dodeca Cell User Manual

Page 17

background image

Problem Cause Solution
5. SDS is precipitating.

a. Running buffer is too cold. a. Set the refrigerated

circulator to maintain a

buffer temperature of

approximately 20 ºC.

6. Agarose slides out

a. The agarose is less

a. Carefully measure the

of the cassette

then 1.0%.

agarose powder to 1.0%.

inside the cell.

7. Buffer in the anode side

a. An electrochemical

a. We recommend using

of the tank turns yellow

reaction is taking place–

fresh buffer for each run.

after a few runs.

a component in the

Tris buffer is oxidized.

This reaction always

occurs during

electrophoresis, but

with the Dodeca cell

the large surface area of

the plate electrodes

intensifies this reaction

and thus causes the

buffer to turn yellow

very quickly.

Section 6
Product Information and Accessories

PROTEAN Plus Dodeca Cell

Catalog

Number

Description

165-4150

PROTEAN Plus Dodeca Cell, 100/120V, includes tank and lid,
buffer recirculation pump with tubing.

165-4151

PROTEAN Plus Dodeca Cell, 220/240V,, includes tank and lid,
buffer recirculation pump with tubing.

165-4153

Replacement Tubing Kit, complete, includes all tubing required to
set up the buffer recirculation pathway.

165-4154

Replacement Gasket Assembly

165-4155

Electrode Card: ANODE

165-4156

Electrode Card: CATHODE

165-4157

Replacement Lid

165-4166

Manifold tubing, qty 11

165-4167

Buffer Exhaust Tubing, includes quick-connect fittings.

165-4158

Buffer Recirculation Pump, 100/120V

165-4159

Buffer Recirculation Pump, 220/240V

165-3320

Gel Releaser Tool, 5

Tygon is a registered trademark of Norton Co.

15