beautypg.com

1 introduction – Bio-Rad PROTEAN Plus Dodeca Cell User Manual

Page 3

background image

Section 1
General Information

1.1 Introduction

The PROTEAN Plus Dodeca Cell* is a multi-cell for polyacrylamide gel electrophoresis.

It is primarily intended to run immobilized pH gradient (IPG) strips for the second dimension
of 2-D electrophoresis. It runs one to twelve large format gels simultaneously. The
PROTEAN Plus Dodeca cell system includes the Buffer Recirculation Pump that aids in
maintaining a constant temperature during the electrophoresis experiment. This system uses
the new PROTEAN Plus hinged spacer plates. These plates are strategically sealed at one
side with a silicon strip to create a hinged cassette. The hinge is required to prevent current
leakage and to achieve optimal results.

Specifications of the PROTEAN Plus Dodeca Cell

Tank and lid

Acrylic

Anode

Platinum-coated titanium plate electrode

Cathode

Stainless steel plate electrode

Drain line

Braided PVC

Drain line connectors

Polypropylene

Cooling core

Ceramic

Cooling core connector tubing

Braided PVC

Manifold tubing

Tygon tubing 1/4" ID (required for PROTEAN II
precast and handcast gels)

Buffer coil

Acrylic

Gaskets

Silicone rubber

Maximum buffer volume

22.5 liters (plate width: 27cm)

Minimum buffer volume

16.8 liters (plate width: 20cm)

Overall size (LxHxW)

Tank and lid

39.5 x 34.5 x 34.5 cm

Pump

19.5 x 12 x 17.5 cm

Precast gel compatibility

PROTEAN II Ready gel precast gels with IPG comb

Handcast gel compatibility

PROTEAN Plus hinged spacer plates
PROTEAN II 20 x 20 cm glass plates

Hinged spacer plate dimensions

26.8 x 22.5 cm; 26.8 x 20.5 cm

Slab gel dimension

25 x 20.5 cm; 20 x 20.5 cm; 18.5 x 20 cm; or

18.3 x 19.3 cm

Power safety limits

1000V, 250W

Weight

17.5 kg

*Patent pending

1