5

Chemical Compatibility
Mini-PROTEAN Tetra cell components are not compatible with

acetone or ethanol. Use of organic solvents voids all warranties.

Call 1-800-4-BIORAD (US) or your local Bio-Rad representative for

technical information regarding chemical compatibility of the

Mini-PROTEAN Tetra cell with various laboratory reagents.

The Mini-PROTEAN are not compatible with repeated exposure to

100% TEMED. Rubbing the combs with TEMED prior to casting

will destroy the structural integrity of the combs over time.

1.4 Safety
Power to the Mini-PROTEAN Tetra cell is supplied by an external

DC voltage power supply (not included). The output of this power

supply must be isolated from external ground to ensure that the

DC voltage output floats with respect to ground. All Bio-Rad power

supplies meet this important safety requirement. Regardless of the

power supply used, the maximum specified operating parameters

for the Mini-PROTEAN Tetra cell are as follows
• 600 V DC maximum voltage limit
• 500 W maximum power limit
• 40°C maximum ambient temperature limit
The current to the cell enters the unit through the lid assembly,

which provides a safety interlock to the user. The current to the

cell is broken when the lid is removed. Always turn off the power

supply before removing the lid. Do not attempt to use the cell

without the safety lid.
Important: This Bio-Rad product is designed and certified to meet

IEC61010-1 and EN61010-1* safety standards. Certified products

are safe to use when operated in accordance with the instruction

manual. This instrument should not be modified or altered in any

way. Alteration of this instrument will
• Void the warranty
• Void the IEC61010-1 and EN61010-1 certifications, and
• Create a potential safety hazard

Bio-Rad is not responsible for any injury or damage caused by

use of this instrument for purposes other than those for which it is

intended or by modifications of the instrument not performed by

Bio-Rad or an authorized agent.

*IEC61010-1 and EN61010-1 are internationally accepted electrical safety

standards for laboratory instruments.