Troubleshooting Guide (cont.)

Problem

Cause

Solution

Poor end well

formation

Incorrect catalyst

formation

Monomer solution

not degassed.

Oxygen inhibits

polymerization

Prepare fresh catalyst solution,

or increase the catalyst

concentration of the stacking

gel to 0.06% APS and 0.12%

TEMED

Degas monomer solution

immediately prior to casting the

stacking gel

Webbing/excess

acrylamide behind

the comb

Incorrect catalyst

concentration

Prepare fresh catalyst solution,

or increase the catalyst

concentration of the stacking

gel to 0.06% APS and 0.12%

TEMED

The pressure cams

on the casting frame

are difficult to close

or make noise when

closed

Powder residue has

built up at the pivot

of the pressure cams

Rinse or wipe off the powder

residue before each use

*Polyacrylamide gels are described by reference to two

characteristics:

1. The total monomer concentration, (%T) and

2. The crosslinking monomer concentration (%C).

g acrylamide + bis-acrylamide x 100%

Total Volume

g bis-acrylamide x 100%

g acrylamide + bis-acrylamide