beautypg.com

Bio-Rad Mini-PROTEAN® Tetra Cell User Manual

Page 14

background image

10

a. Set the clamping frame to the open position on a clean flat

surface (see Figure 4a).

b. Place the first gel sandwich or gel cassette (with the short

plate facing inward) onto the gel supports; gel supports are

molded into the bottom of the clamping frame assembly;

there are two supports in each side of the assembly. Note

that the gel will now rest at a 30° angle, tilting away from the

center of the clamping frame.Please use caution when

placing the first gel, making sure that the clamping frame

remains balanced and does not tip over. Now, place the

second gel on the other side of the clamping frame, again by

resting the gel onto the supports. At this point there will be

two gels resting at an angle, one on either side of the

clamping frame, tilting away from the center of the frame (see

Figure 4b).

Note: It is critical that gel cassettes are placed into the clamping

frame with the short plate facing inward. Also, the clamping

frame requires 2 gels to create a functioning assembly. If an odd

number of gels (1 or 3) is being run, you must use the buffer dam

(see Figure 4b).

c. Using one hand, gently pull both gels towards each other,

making sure that they rest firmly and squarely against the

green gaskets that are built into the clamping frame; make

certain that the short plates sit just below the notch at the

top of the green gasket.

d. While gently squeezing the gel sandwiches or cassettes

against the green gaskets with one hand (keeping constant

pressure and both gels firmly held in place), slide the green

arms of the clamping frame over the gels, locking them

into place. Alternatively, you may choose to pickup the

entire assembly with both hands, making sure that the

gels do not shift, and simultaneously sliding both arms of

the clamping frame into place (see Figure 4c).

The arms of the clamping frame push the short plates of each

gel cassette up against the notch in the green gasket, creating a

leak-proof seal (check again to make certain that the short plates

sit just below the notch at the top of the green gasket). At this

point, the sample wells can be washedout with running buffer,

and sample can be loaded (Figure 4d).

Note: If running more than 2 gels, repeat steps 1a–d with the

companion running module