beautypg.com

Bio-Rad Mini-PROTEAN® Tetra Cell User Manual

Page 13

background image

9

Ready Gel

®

Precast Gels

1. Ready Gel Cassette Preparation

Note: The Mini-PROTEAN Tetra cell is guaranteed for use with

Bio-Rad’s Ready Gel and Mini-PROTEAN

®

precast gels. For

more information, go to www.bio-rad.com/mpgels.

a. Remove the Ready Gel from the storage pouch.

b. Gently remove the comb and rinse the wells thoroughly with

distilled water or running buffer.

c. Cut along the dotted line at the bottom of the Ready Gel

cassette with a razor blade.

d. Pull the clear tape at the bottom of the Ready Gel cassette

to expose the bottom edge of the gel.

e. Repeat for second Ready Gel.

Note: If only one or three gels are to be run, use the mini cell

buffer dam.

2.2 Electrophoresis Module Assembly and Sample Loading

Required materials:

• Clean and dry Mini-PROTEAN Tetra cell tank

• Electrophoresis module (electrode assembly module only for

1 or 2 gels; for 3 or 4 gels also use the companion running

module)

• Running buffer (700 ml for 2 gels; 1000 ml for 4 gels)

• Ready Gel precast gels or hand-cast gels

• PowerPac™ Basic power supply

1. Assembly

Note: When running 2 gels only, use the electrode assembly (the

one with the banana plugs), not the companion running module

(the one without the banana plugs). When running 4 gels, both

the electrode assembly and the companion running module

must be used, for a total of 4 gels (2 gels per assembly).