Programming – ADS Environmental Services Enigma-hyQ QR 775017 A2 User Manual
Page 17

ADS Enigma/Enigma-Compact/Enigma-hyQ User Manual
13
Note: Hydrophone loggers cannot be inserted into the case with
hydrophones connected. Reconnect the hydrophone to the logger after
programming and before the first epoch.
The following table provides instructions and descriptions for the controls
involving programming and readback:
Control
Use/Indication
Description
Green
Button
Short Press
Reports status of last case programming
operation
Green
Button
Long Press
Starts programming sequence
Red Button
Short Press
Reports status of last case readback
operation
Red Button
Long Press
Starts readback sequence
LEDs (8)
On
Indicates success of appropriate
programming/ readback
LEDs (8)
Flash
Indicates logger is detected but failed to
program/readback
LEDs (8)
Off
Indicates no logger is detected
Note: The Enigma–Compact has 3 loggers and LEDs.
Programming
Program the loggers using the case in the following way:
1. Make sure the loggers to be programmed are in the case.
2. Press and hold the green button for about 1 second. Check for the
“walking LEDs” and fast beeping, and then close the case. The
beeping should slow down to about 1 beep per second after
approximately 8 seconds. This indicates that the case is programming
the loggers.
3. After hearing a long beep, open the case.
4. Verify whether the LED for each logger is on, indicating successful
programming. Please note that logger status is displayed for only 8
seconds. To re-display status information, follow the instructions in
the following section, Checking Status - Programming.
If the case fails to program a particular logger, clean the top of the logger and
the appropriate optical window in the top of the case and then repeat the
programming procedure.