beautypg.com

Calibrate, Example calibration screen, Calibrate speed sensor – Great Plains VARIABLE RATE KIT Operator Manual User Manual

Page 18: Calibrate material, Pm near field speed at step 92

background image

18

Air Drill Variable Rate Kit

Front

Parts

Great Plains Mfg., Inc.

166-263M

Front

Parts

6/23/2010

Calibrate

Calibrate Speed Sensor

88. Perform a Calibrate Ground Speed.

Calibrate Material

Perform a calibration. The procedure is similar to that for
crank-set operations.

89. Consult the Seed Rate manual (167-085B) for (the

approximate) manually-set rate for your material.

90. Check that final Range gear pairing is the same on:

• the chart,
• the meter, and
• the Meter gear Range in the Channel Setup.

91. Enter the calibration screen.

At this time, the linear actuator becomes active. The
scale indicator moves to approximately mid-scale,
then stops.

Pointing to a specific value is not required, but
needs to be in the range 30

° to 95°. Great Plains

recommends using a scale setting that is close to
your expected target rate.

Use the Inc+/Dec- softkeys on the monitor console
to adjust the indicator to the target rate, or at least to
within the 30

°-95° range.

92. Manually crank the meter for at least the number of

turns shown in the instruction of the Seed Rate
Manual for

1

10

ha or

1

10

ac.

The exact number of revolutions, cranking rate

b

,

and precise starting and stopping handle angles are
not critical, as the system reads meter revolutions
accurately, and can compensate for shaft speed,
seed size and partial turns.

What matters is getting a large sample, to reduce
errors and increase confidence in the calibration.

Note: By calibrating at the target rate, and for

1

10

ha, you

establish a comfort level that the drill is set up cor-
rectly, in particular, that you are in the correct gear
Range for the desired application rate.

93. Weigh the dispensed material and enter it in the

menu. The monitor computes a Calibration Con-
stant for this material. Save this result.

b. If no “Variable Cal Const” was selected at step 84, crank at for most accurate results:

ADC2350/B, NTA907/3007HD: 2 to 2

1

2

revolutions per second

NTA607/2007HD: 1 to 1

3

4

revolutions per second

Example Calibration Screen:

Parameter

Value

Comment

CHANNEL

1

Example

material

Wheat HRW

Example

Density

60.0

LB/BU

Example

Calibration
Constant

77591

PUL/FT

From chart

Target Meter
rpm

20

RPM

Example

# Meter
Revs

30

REV

Example

Pulse Count 0

PUL

Pre-Cal.

New Calib
Const

______

PUL/FT

Total #
Towers

5

Per Imple-
ment

Amount
Dispensed

______

LBS

From Scale

Example:
ADC2350B cart and NTA3510 drill
Wheat, 2-star meter, single seed hopper in use
High Range at gearbox
Target rate of 133 kg/ha would be a scale setting of:

50

° (from Seed Rate manual charts)

A sample size of

1

10

ha requires crank:

77 turns (from Seed Rate manual instructions)