Gossen Starlite User Manual

Page 32

background image

32

GOSSEN Foto- und Lichtmeßtechnik

10.3 Measuring the luminance candela / m

2

(cd/m

2

) or footLambert (fL)

– Set the measuring head to reflected light

measuring –1°

or 5°

. The luminance

function is set and shown in the display.

– Focus on the subject to be measured via the

viewfinder

– Take a measurement using the measuring

button

The selected display unit and measured
luminance are displayed.

10.4 Measuring time-integral values

(lxs, fcs, cds/m

2

, fLs)

– Set measuring head to reflected light

or incident light measuring.
The corresponding display unit is set
and shown in the display.

– Set with the right function button

the

function flash light

.

– Pre-select the desired measuring time using

the setting wheel; this can be found on the right
hand side of the display panel. Activated half
time stop increments are also displayed.

– Start flash measuring using the measuring but-

ton

– cord/noncord (chapter 6.1, page 23).

– The measured value, which is calculated to

1 second, is shown in the pre-selected display
unit.

By preselecting the measuring time, the ambient
light part is correspondingly taken into account.