10.3 Measuring the luminance candela / m
2
(cd/m
2
) or footLambert (fL)
– Set the measuring head to reflected light
measuring –1°
or 5°
. The luminance
function is set and shown in the display.
– Focus on the subject to be measured via the
viewfinder
– Take a measurement using the measuring
button
The selected display unit and measured
luminance are displayed.
10.4 Measuring time-integral values
(lxs, fcs, cds/m
2
, fLs)
– Set measuring head to reflected light
or incident light measuring.
The corresponding display unit is set
and shown in the display.
– Set with the right function button
the
function flash light
.
– Pre-select the desired measuring time using
the setting wheel; this can be found on the right
hand side of the display panel. Activated half
time stop increments are also displayed.
– Start flash measuring using the measuring but-
ton
– cord/noncord (chapter 6.1, page 23).
– The measured value, which is calculated to
1 second, is shown in the pre-selected display
unit.
By preselecting the measuring time, the ambient
light part is correspondingly taken into account.