5.5 High pressure pump

■■■■■■■■■■■■■■■■■■■■■■

78

■■■■■■■■

850 Professional IC Anion – MSM-HC

1

2

Figure 40

Inserting the piston seal into the pump head

4 Replacing the piston cartridge

Screw the assembled piston cartridge back into the pump head and
tighten, first by hand, then additionally by approx. 15° with a
wrench.

Cleaning the inlet valve and outlet valve

1 Removing valves

Unscrew the connection capillary for the auxiliary piston (18-1)
from the outlet valve holder.

Unscrew the holders for the inlet and outlet valves and remove
valves.