6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
102
■■■■■■■■
850 Professional IC Anion – MSM-HC
Problem
Cause
Remedy
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Connections – dead vol-
ume in the system.
Check the connections (see Chapter 3.5, page
18) (use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).
Guard column – diminished
performance.
■
Replace guard column (see Chapter 3.20,
page 62).
Separation column – dead
volume at column head.
■
Install separation column in reverse flow
direction (if permitted by the information
sheet).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65).
Greater rise in the
baseline
MSM-HC – reduced
capacity.
Regenerate the MSM-HC (see Chapter
5.11.3.1, page 88).
Precision problems -
significant scatter-
ing of the measured
values
Sample – Gas bubbles in
the sample.
Use sample degasser (see Chapter 3.13, page
44).
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 3.14.1, page 45).
Sample – Rinsing volume
too low.
Prolong the rinsing time (see Chapter 5.8,
page 85).
Injection valve – defective.
Request Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.14.4, page 97).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65).
Eluent – Gas bubbles in the
eluent.
Check connections of the eluent degasser (see
Chapter 3.9, page 35).
High pressure pump –
defective.
Request Metrohm Service.
Vacuum is not being
built
Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.
■
Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module: