6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

850 Professional IC Anion – MSM-HC

Problem

Cause

Remedy

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Connections – dead vol-
ume in the system.

Check the connections (see Chapter 3.5, page
18)
(use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 3.20,
page 62)
.

Separation column – dead
volume at column head.

Install separation column in reverse flow
direction (if permitted by the information
sheet).

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65)
.

Greater rise in the
baseline

MSM-HC – reduced
capacity.

Regenerate the MSM-HC (see Chapter
5.11.3.1, page 88)
.

Precision problems -
significant scatter-
ing of the measured
values

Sample – Gas bubbles in
the sample.

Use sample degasser (see Chapter 3.13, page
44)
.

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.14.1, page 45)
.

Sample – Rinsing volume
too low.

Prolong the rinsing time (see Chapter 5.8,
page 85)
.

Injection valve – defective.

Request Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.14.4, page 97)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65)
.

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 3.9, page 35)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module: