19 connecting the instrument, 1 connecting the instrument to the pc, 2 connecting the instrument to mains supply – Metrohm 850 Professional IC Anion MSM-HC User Manual
Page 71: Connecting the instrument to the pc, Connecting the instrument to mains supply, Figure 34, Connection detector – msm-hc, Screws (34-4)
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
850 Professional IC Anion – MSM-HC
■■■■■■■■
61
1
2
3
4
Figure 34
Connection Detector – MSM-HC
1
Detector inlet capillary
2
MSM-HC outlet capillary
Labeled with Detector.
3
Coupling 6.2744.040
4
Pressure screws, short 6.2744.070
3.19
Connecting the instrument
3.19.1
Connecting the instrument to the PC
Note
The instrument must be turned off when connecting the PC.
1 Connecting the USB cable
Connect the PC connection socket (2-18) of the instrument to a USB
connector of the computer via the 6.2151.020 USB cable.
3.19.2
Connecting the instrument to mains supply
Warning
The power supply unit must not get wet. Protect it against the direct
effect of liquids.
Mains cable
Which mains cable is supplied depends on the location:
■
6.2122.020 with plug SEV 12 (Switzerland, …)
■
6.2122.040 with plug CEE(7), VII (Germany, …)
■
6.2122.070 with plug NEMA 5-15 (USA, …)
The mains cable is three-core and provided with a plug with grounding. If
another plug has to be mounted, the yellow/green conductor (IEC stand-
ard) must be connected to the protective ground (protection class I).