13 conductivity measuring system – Metrohm 850 Professional IC Anion MSM-HC User Manual
Page 118
7.13 Conductivity measuring system
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
850 Professional IC Anion – MSM-HC
7.13
Conductivity measuring system
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm without range switching
Noise
< 0.1 nS at 1 µS/cm
Deviations from
the linearity
■
< 0.1 % for conductivity values greater than 16 µS/cm (typical for
analyses with MSM)
Drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 0.3 nS/cm typical for chemical suppression
Conductivity
detector
Cell volume
0.8 µL
Cell constant
■
Individual calibration data saved in the detector
■
adjustable range: 13.0…21.0 /cm
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)
Cell tempera-
ture
20…50 °C in increments of 5 °C
Temperature
stability
< 0.001 °C
Temperature
compensation
0…5 %/K adjustable, default 2.3 %/K
Heating up time
< 30 minutes (40 °C)