6 Troubleshooting

850 Professional IC Anion – MSM-HC

■■■■■■■■

101

Problem

Cause

Remedy

MSM-HC – regeneration or
rinsing solution flow prob-
lems.

Check the flow of the regeneration and rinsing
solution (see Chapter 3.16.2, page 51).

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean contact surfaces of the column chip
(with alcohol).

Column chip defective.

1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.

Individual peaks
greater than expec-
ted

Sample – Carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

MSM-HC – No (or
insufficient) pump-
ing of regeneration
or rinsing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure .

Peristaltic pump – filter
blocked .

Replace the filter .

MSM-HC – backpressure
too high.

Clean the MSM-HC (see Chapter 5.11.3.2,
page 89)
or replace parts
(see Chapter
5.11.3.3, page 92)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing .

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.14.4, page 97)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (33-

1

).

Switch the instrument off and on again
(after 15 seconds).