6.2 Problems
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
pH Meter/Conductometer
6.2
Problems
The following list describes some general problems that might occur dur-
ing measurements. Furthermore, the possible causes and solution
approaches are described.
NOTE
Sensor treatment
Follow the instructions given in the respective leaflets for sensors clean-
ing and maintenance.
6.2.1
Troubleshooting
Problem
Cause
Remedy
Measured value set-
ting is sluggish.
The glass membrane or the
diaphragm is contami-
nated.
■
Clean the electrode following the instruc-
tions in the leaflet.
No measuring sig-
nal.
The sensor is not con-
nected.
■
Connect the sensor.
Wrong measuring channel
is selected.
■
Select the correct measuring channel.
The sensor is defective.
■
Replace the sensor.
The cable is defective.
■
Replace the cable.
The electrode's reference
system contains air.
■
Perform an electrode maintenance as
described in the leaflet.
The measuring input and/
or the measuring channel
is defective.
■
Send the measuring instrument to the
Metrohm Service for inspection and, if nec-
essary, repair.
The instrument does
not start.
The instrument battery is
not charged.
■
Connect the instrument to the power sup-
ply unit to charge it.
The battery is only charged when the
instrument is on.
(total charging time: approx. 9 hours)
The measured value
drift criterion is not
fulfilled.
The glass membrane or the
diaphragm is contami-
nated.
■
Clean the electrode following the instruc-
tions in the leaflet.