Metrohm 914 pH / Conductometer User Manual
Page 55
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
pH Meter/Conductometer
■■■■■■■■
45
■
on
■
off: is the default setting
pH
Slope lower limit
Editing dialog for the entry of the lower limit value.
■
Default value: 95.0 / input range: 1.0 - 999.9%
pH
Slope upper
limit
Editing dialog for the entry of the upper limit value.
■
Default value: 103.0 / input range: 1.0 - 999.9%
pH
Limit value
pH(0)
Selection dialog for selecting whether the limit value is to be applied.
■
on
■
off: is the default setting
pH
pH(0) lower limit
Editing dialog for the entry of the lower limit value.
■
Default value: 6.40 / input range: 0.00 - 99.99
pH
pH(0) upper
limit
Editing dialog for the entry of the upper limit value.
■
Default value: 8.00 / input range: 0.00 - 99.99
K
Limit value c
Selection dialog for selecting whether the limit value is to be applied.
■
on
■
off: is the default setting
K
c lower limit
Editing dialog for the entry of the lower limit value.
■
Default value: 0.400 /cm / input range: 0.001 - 500 /cm
K
c upper limit
Editing dialog for the entry of the upper limit value.
■
Default value: 0.550 /cm / input range: 0.001 - 500 /cm
New sensor
▶
Menu dialog with the individual menu lines for entering a new sensor.
Sensor type
Selection dialog for the sensor type.
■
pH electrode
■
Conductivity
■
Metal electrode
■
Temp. sensor
■
Other sensor...
Sensor name
Editing dialog for entering the sensor name.
Order number
Editing dialog for entering the Order number.
Serial number
Editing dialog for entering the Serial number.
pH
Slope
Editing dialog for the entry/modification of the Slope.