4 sensors, Sensors – Metrohm 914 pH / Conductometer User Manual
Page 52
4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
pH Meter/Conductometer
■
drift-dependent
The measured value will be applied when the value is stable according to the
drift criterion.
The drift thresholds are preset and cannot be changed:
– pH measurement: 0.028 pH/min
– Potential measurement U/mV: 1.875 mV/min
– Temperature measurement T/°C: 0.974 °C/min
For the conductivity, various drift thresholds are stored (depending on the
measuring range):
– 0.005 mS/cm/min up to 16 µS/cm
– 0.5 mS/cm/min between 16 µS/cm and 1 mS/cm
– 10 mS/cm/min for 1 mS/cm and higher
For instruments with two measuring channels, the primary measuring chan-
nel for the fulfillment of the drift criterion has to be selected.
– Primary channel
Selection dialog for selecting the measuring channel that has to fulfill
the drift criterion.
pH/mV or Cond.
Output date/time
Selection dialog for selecting whether a time stamp is to be assigned to a mea-
sured value.
■
on: Measured values contain a time stamp on the report.
■
off: Measured values do not contain a time stamp on the report.
Default value: off
Output headers
Selection dialog for selecting how the headers are output.
■
once
■
always: is the default setting
■
off
Calibration data
Selection dialog for selecting whether the main Calibration data is assigned for
the output of the measured values.
■
on
■
off: is the default setting
4.7.4
Sensors
The Sensors menu dialog is shown below with the structure and the
description.