7.1 Coulometric Karl Fischer titrations (KFC)
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
899 Coulometer
Selection
on | off
Default value
on
7.1.3
Control parameters
Menu
▶ Parameters ▶ Control parameters
The control parameters for the endpoint are defined under Control
parameters.
Endpoint at
Measured value for the endpoint.
Input range
–1250.0 ... 1250.0 mV
Default value
50.0 mV
Selection
off
Note
Note that the parameter Endpoint at is linked with the parameters
Dynamics and I(pol). Explanations in this connection can be found in
chapter 10.4 Control parameters and polarization current, page 95.
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection
slow | optimal | fast | user
Default value
optimal
slow
For samples with a low water content or samples which release their
moisture only slowly.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For uncritical samples with high water content.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 3, page
77.
Dynamics
This parameter can only be edited with Titration rate = user.