7 Parameters
899 Coulometer
■■■■■■■■
75
Selection
off
Default value
off
Measured value display
If this parameter is activated, then the currently measured value is dis-
played during the conditioning.
Selection
on | off
Default value
off
7.1.2
Start conditions
Menu
▶ Parameters ▶ Start conditions
The parameters that are carried out before the start of titration are
defined under Start conditions.
Pause
Waiting time, for example for dissolving the sample. During this time no
iodine is generated.
Input range
0 ... 999999 s
Default value
0 s
Request sample ID
Selection of the sample identification that is queried in the method run.
Selection
ID1 | ID2 | ID1&ID2 | off
Default value
off
Request sample size
If this parameter is activated, then the value for the sample size will be
requested.
Selection
on | off
Default value
on
Request sample unit
If this parameter is activated, then the unit for the sample size will be
requested.
Selection
on | off
Default value
off
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is deactivated, then the titration will be started in
the background.