beautypg.com

3 formula editor, Formula editor – Metrohm 899 Coulometer User Manual

Page 49

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation

899 Coulometer

■■■■■■■■

39

5.3

Formula editor

The formulas for the calculations are entered with the formula editor. The
formula editor is equipped with an automatic syntax check. This is trig-
gered as soon as a formula is applied. The generally valid rules of priority
apply for the calculation operations.

Variable

Description

C00

Sample size

EP#

Water quantity at the endpoint EP# (# = 1…9)

CI#

Sample identification (# = 1…2)

R#

Result (# = 1…5)

CV0#

Common variable (# = 1…5)

SMN#

Mean value of result R# (# = 1…5)

Var

List of additional variables (see "Variables", page 39)

"#" stands for a sequential number that you must enter manually. Exam-
ple: if you apply the variable EP# in the formula, only EP is entered. You
will still need to enter the number yourself.

The meanings of the editing functions are explained in chapter 5.2.4,
page 37
.

Variables
Pressing Var displays a list with additional variables. You can enter these
variables either directly into the formula or also by selecting them from the
list and applying them with [OK].

Variable

Description

MCQ

End quantity, i.e. total amount of removed water at the
end of the titration (in µg)

MCD

Duration of the entire titration

MDC

Drift for drift correction