9 Troubleshooting
899 Coulometer
■■■■■■■■
91
Problem
Cause
Remedy
The parameter Min. rate
is set too low.
■
Set the parameter Min. rate higher.
■
Reset the Control parameters to default
values.
The results are
spread widely.
The sample is nonhomoge-
nous.
■
Homogenize the sample if possible.
■
Possibly use a larger sample size.
The sample addition is
poorly reproducible.
■
Observe the instructions in chapter 4.3,
page 31.
The drift fluctuates.
■
Ensure stable drift.
9.2
Miscellaneous
Problem
Cause
Remedy
No report is printed.
The printer is not recog-
nized by the instrument.
■
Switch the 899 Coulometer off and then
back on again.
■
Use a hub on the USB connector and con-
nect the printer to the hub.
The printer model is not
compatible.
Use a printer which fulfills the required specifi-
cations (see Chapter 10.3.4, page 95).
The USB keyboard
or the PC mouse
does not function.
The keyboard or mouse is
not recognized by the
instrument.
■
Switch the 899 Coulometer off and then
back on again.
■
Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.
The keyboard or the mouse
is not compatible.
Use a model which fulfills the required specifi-
cations (see Chapter 10.3, page 93).
The display is no
longer readable.
The contrast is set incor-
rectly.
Adjust the contrast correctly (see Chapter 6.1,
page 62).
Mettler XP balances
send "R" or "O ----"
as ID1.
The automatic calibration
of the balance is activated.
Deactivate the automatic calibration.
Message 020-511
"Action not possi-
ble" is displayed.
The USB flash drive is no
longer connected.
1. Connect the USB flash drive.
2. Switch the instrument off and then back
on again.
The USB flash drive is full.
■
Use a different USB flash drive.
■
Delete files with the aid of a PC.