9 Troubleshooting

899 Coulometer

■■■■■■■■

91

Problem

Cause

Remedy

The parameter Min. rate
is set too low.

Set the parameter Min. rate higher.

Reset the Control parameters to default
values.

The results are
spread widely.

The sample is nonhomoge-
nous.

Homogenize the sample if possible.

Possibly use a larger sample size.

The sample addition is
poorly reproducible.

Observe the instructions in chapter 4.3,
page 31
.

The drift fluctuates.

Ensure stable drift.

9.2

Miscellaneous

Problem

Cause

Remedy

No report is printed.

The printer is not recog-
nized by the instrument.

Switch the 899 Coulometer off and then
back on again.

Use a hub on the USB connector and con-
nect the printer to the hub.

The printer model is not
compatible.

Use a printer which fulfills the required specifi-
cations (see Chapter 10.3.4, page 95).

The USB keyboard
or the PC mouse
does not function.

The keyboard or mouse is
not recognized by the
instrument.

Switch the 899 Coulometer off and then
back on again.

Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.

The keyboard or the mouse
is not compatible.

Use a model which fulfills the required specifi-
cations (see Chapter 10.3, page 93).

The display is no
longer readable.

The contrast is set incor-
rectly.

Adjust the contrast correctly (see Chapter 6.1,
page 62)
.

Mettler XP balances
send "R" or "O ----"
as ID1.

The automatic calibration
of the balance is activated.

Deactivate the automatic calibration.

Message 020-511
"Action not possi-
ble" is displayed.

The USB flash drive is no
longer connected.

1. Connect the USB flash drive.
2. Switch the instrument off and then back

on again.

The USB flash drive is full.

Use a different USB flash drive.

Delete files with the aid of a PC.