7 safety specifications, 8 electromagnetic compatibility (emc), 9 ambient temperature – Metrohm 858 Professional Sample Processor User Manual
Page 58: Safety specifications, Electromagnetic compatibility (emc), Ambient temperature

7.7 Safety specifications
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
858 Professional Sample Processor
7.7
Safety specifications
Design and testing
According to EN/IEC/UL 61010-1, EN/IEC 61010-2-081, CSA-C22.2 No.
61010-1, Protection Class
Ⅰ
Safety instructions
This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.
7.8
Electromagnetic compatibility (EMC)
Emission
Standards fulfil-
led
■
EN/IEC 61326
■
EN 55022 / CISPR 22
■
EN/IEC 61000-3-2
Immunity
Standards fulfil-
led
■
EN/IEC 61326
■
EN/IEC 61000-4-2
■
EN/IEC 61000-4-3
■
EN/IEC 61000-4-4
■
EN/IEC 61000-4-5
■
EN/IEC 61000-4-6
■
EN/IEC 61000-4-8
■
EN/IEC 61000-4-11
■
EN/IEC 61000-4-14
■
NAMUR
7.9
Ambient temperature
Nominal working
range
5…45 °C
Relative humidity <80% (below 30 °C)
Relative humidity <50% (below 45 °C)
Storage
–20…60 °C
Relative humidity <95% (below 40 °C)
Relative humidity <85% (below 50 °C)
Relative humidity <50% (below 60 °C)
Transport
–40…60°C
Relative humidity <95% (below 40 °C)
Relative humidity <85% (below 50 °C)
Relative humidity <50% (below 60 °C)