Page 45: Connecting a dosing device

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
858 Professional Sample Processor
■■■■■■■■
37
Warning
If a Dosino is connected to the 858 Professional Sample Processor then
the connection cable must be equipped with a T.2400.102 ferrite core.
The ferrite core reduces any interference voltages that may occur and
thus ensures compliance with strict EMC standards pursuant to applica-
ble technical norms, see Chapter "Technical Data".
Proceed as follows:
1 Mounting ferrite core
Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.
2 Connect a dosing device
■
Exit the control software.
■
Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.
■
Start the control software.
USB 1
Contr.
MSB 2
MSB 3
T.2400.102
Figure 35
Connecting a dosing device
3.14.2
Connecting a stirrer or titration stand
You can use a magnetic stirrer 801 Stirrer or 803 Ti Stand (stirring "from
below") or the 804 Ti Stand with a rod stirrer 802 Stirrer (stirring "from
above").
Connect a stirrer or a titration stand as follows:
1 Connect a stirrer or titration stand
■
Exit the control software.