Selecting the pump tubing
The pump tubing differs in material, diameter and hence also pumping
capacity. Depending on the application, different pump tubings are used.

The following table provides information on the properties and use of the
pump tubing:

Table 2

Pump tubing

Order
number

Name

Material

Inner
diameter

Use

6.1826.020

Pump tubing (blue/
blue), 2-stopper

PVC (Tygon
ST)

1.65 mm

Pump tubing for online IC
instruments and automa-
tion in voltammetry

6.1826.310

Pump tubing LFL
(orange/green), 3-
stopper

PVC (Tygon)

0.38 mm

Pump tubing for bromate
determination using the
triiodide method.

6.1826.320

Pump tubing LFL
(orange/yellow), 3-
stopper

PVC (Tygon)

0.48 mm

For suppressor solutions,
acceptor solutions for
inline dialysis and for inline
ultrafiltration.

6.1826.330

Pump tubing LFL
(orange/white), 3-
stopper

PVC (Tygon)

0.64 mm

No special applications.

6.1826.340

Pump tubing LFL
(black/black), 3-stop-
per

PVC (Tygon)

0.76 mm

For the sample solution in
inline dialysis.

6.1826.360

Pump tubing LFL
(white/white), 3-stop-
per

PVC (Tygon)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL
(gray/gray), 3-stopper

PVC (Tygon)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yel-
low/yellow), 3-stop-
per

PVC (Tygon)

1.37 mm

For the sample solution in
inline ultrafiltration.