Metrohm 858 Professional Sample Processor User Manual
Page 51

■■■■■■■■■■■■■■■■■■■■■■
4 Handling and maintenance
858 Professional Sample Processor
■■■■■■■■
43
Selecting the pump tubing
The pump tubing differs in material, diameter and hence also pumping
capacity. Depending on the application, different pump tubings are used.
The following table provides information on the properties and use of the
pump tubing:
Table 2
Pump tubing
Order
number
Name
Material
Inner
diameter
Use
6.1826.020
Pump tubing (blue/
blue), 2-stopper
PVC (Tygon
ST)
1.65 mm
Pump tubing for online IC
instruments and automa-
tion in voltammetry
6.1826.310
Pump tubing LFL
(orange/green), 3-
stopper
PVC (Tygon)
0.38 mm
Pump tubing for bromate
determination using the
triiodide method.
6.1826.320
Pump tubing LFL
(orange/yellow), 3-
stopper
PVC (Tygon)
0.48 mm
For suppressor solutions,
acceptor solutions for
inline dialysis and for inline
ultrafiltration.
6.1826.330
Pump tubing LFL
(orange/white), 3-
stopper
PVC (Tygon)
0.64 mm
No special applications.
6.1826.340
Pump tubing LFL
(black/black), 3-stop-
per
PVC (Tygon)
0.76 mm
For the sample solution in
inline dialysis.
6.1826.360
Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon)
1.25 mm
For inline sample dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon)
1.37 mm
For the sample solution in
inline ultrafiltration.