5 Troubleshooting
858 Professional Sample Processor
■■■■■■■■
45
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
Marked rise in pres-
sure
Injection valve – valve
blocked.
Rinse the valve with water in the opposite
direction or have it cleaned (by Metrohm serv-
ice technician).
The Swing Head
either misses the
rack positions
totally or is inaccu-
rate
Sample Processor – The
Swing Head is not correctly
configured.
Enter the correct values for Swing radius,
Swing offset etc. in the control software
under "Configuration".
Sample Processor – The
axial distance is not cor-
rectly configured.
Enter the correct value for Axial distance in
the control software under "Configuration".
Sample Processor – The
wrong rack table is being
used.
Initialize the rack using the function Initialize
rack in the "Manual Control" of the control
software.
Peristaltic pump –
insufficient or no
delivery rate
Peristaltic pump – contact
pressure too weak.
Correctly set the contact pressure .
Peristaltic pump – filter
blocked.
Replace the filter .
Peristaltic pump – pump
tubing defective.
Replace the pump tubing .
Precision problems -
significant scatter-
ing of the measured
values
Injection valve – sample
loop.
Check the installation of the sample loop.
Injection valve – defective.
Contact the Metrohm Service.