4 Operation and maintenance

850 Professional IC –Anion – MCS – Gradient

■■■■■■■■

89

1

2

3

4

5

Figure 46

MSM – Components

1

Union nut

2

MSM connecting piece 6.2832.010

3

MSM rotor A 6.2832.000

4

MSM housing

5

Slot in the MSM housing

Cleaning the MSM

Clean the MSM as follows:

1 Disconnecting the MSM from the IC system

Switch off the instrument.

Disconnect the MSM from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM

Unscrew union nut (46-1) from the MSM housing (46-4).

Pull MSM connecting piece (46-2) and MSM rotor A (46-3) out of
the MSM housing (46-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (46-5) in the MSM housing,
and pull out MSM rotor A (46-3) in this way.

Detach the MSM connecting piece (46-2) from MSM rotor A
(46-3)
.

3 Cleaning the supply and discharge lines

Connect in turn each of the 6 capillary tubings fastened on the
MSM connecting piece (46-2) on the high pressure pump (see
Chapter 2.10, page 31)
and pump through ultra pure water.