4.12 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
94
■■■■■■■■
850 Professional IC –Anion – MCS – Gradient
Order
number
Name
Material
Inner
diameter
Use
6.1826.330
Pump tubing LFL
(orange/white), 3-
stopper
PVC (Tygon)
0.64 mm
No special applications.
6.1826.340
Pump tubing LFL
(black/black), 3-stop-
per
PVC (Tygon)
0.76 mm
For the sample solution in
inline dialysis.
6.1826.360
Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon)
1.25 mm
For inline sample dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon)
1.37 mm
For the sample solution in
inline ultrafiltration.
4.12.2.2
Pump tubing connection with filter
The 6.2821.130 filters (47-2) should be changed every 3 months, more
frequently at higher backpressure.
1
2
3
Figure 47
Pump tubing connection – Changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter housing
Replacing the filter
1 Unscrewing filter screw
■
Screw the filter screw (47-3) out of the tubing olive (47-1) with
the aid of two 6.2621.000 adjustable wrenches.
2 Replacing the filter
■
Remove the old filter (47-2) with tweezers.
■
Place the new filter (47-2) flat in the tubing olive (47-1) with
tweezers.