beautypg.com

Metrohm 850 Professional IC Anion MCS Gradient User Manual

Page 29

background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

850 Professional IC –Anion – MCS – Gradient

■■■■■■■■

19

Note

Up to two detectors can be positioned and connected.

Caution

The instrument must be turned off when connecting a detector.

1 Removing rear panel

Unscrew knurled screws (6-1) on the rear panel.

Remove handle if still fastened to the instrument.

Remove rear panel (6-2).

2 Positioning the detector

Put the detector through this opening and position it on the sup-
port surface intended for this purpose. Slide it right up to the
front.

3 Replacing rear panel

Insert the detector cable in a cable feed-through (6-3) on the rear
panel (6-2).

Place the detector output capillary in a suitable capillary feed-
through.

Replace (6-2) rear panel.
(Optionally, the handle can be remounted higher up and used as
a holder for MPaks.)

Tighten (6-1) knurled screws.

4 Connecting the detector

Note

The instrument has two detector connection sockets (6-5), Detec-
tor 1
and Detector 2. You must ensure that the selected connec-
tor corresponds to the connector entered in the MagIC Net
method.

Recommendation: Use Detector 1 as standard. In the AnCat sys-
tem with 2 detectors: Detector 1 for anions, Detector 2 for cati-
ons.