beautypg.com

Figure 30, Install pump tubing connection without filter – Metrohm 850 Professional IC Anion MCS Gradient User Manual

Page 62

background image

2.17 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

850 Professional IC –Anion – MCS – Gradient

1

2

3

Figure 30

Install pump tubing connection without filter

1

Union nut

2

Adapter

3

Tubing olive

Slide union nut (30-1) onto the pump tubing.

Select a suitable adapter (30-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 52)
.

Place the tubing olive (30-3) onto the pump tubing.

Screw the union nut (30-1) onto the tubing olive (30-3).

4 Inserting the pump tubing

Press the contact pressure lever all the way down.

Place the pump tubing in the tubing cartridge. The stoppers
(28-3) must snap into the corresponding holders of the tubing
cartridge.

5 Inserting the tubing cartridge

Hang the tubing cartridge in the mounting pin and press in the
cartridge holder until the snap-action lever snaps in.

6 Connecting the capillaries

Screw the respective capillaries tightly to the two tubing olives
with PEEK pressure screws (28-1).

Table 1

Pump tubings and suitable adapters

Pump tubing

Adapter

6.1826.020 (blue/blue)

6.1826.310 (orange/green)

6.1826.320 (orange/yellow)

6.1826.330 (orange/white)