beautypg.com

Metrohm 850 Professional IC Anion MCS Gradient User Manual

Page 112

background image

5.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

850 Professional IC –Anion – MCS – Gradient

Problem

Cause

Remedy

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 64)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (34-

1

).

Turn the instrument off and on again (after
15 seconds).

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Connections – dead vol-
ume in the system.

Check the connections (see Chapter 2.5, page
13)
(use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 2.21,
page 61)
.

Separation column – dead
volume at column head.

Install separation column in reverse flow
direction (if permitted by the information
sheet).

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 64)
.

Greater rise in the
baseline

MSM – reduced capacity.

Regenerate the MSM (see Chapter 4.11.3.1,
page 87)
.

Precision problems -
significant scatter-
ing of the measured
values

Sample – Gas bubbles in
the sample.

Use sample degasser (see Chapter 2.13, page
39)
.

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 2.14.1, page 40)
.

Sample – Rinsing volume
too low.

Prolong the rinsing time (see Chapter 4.8,
page 84)
.

Injection valve – defective.

Request Metrohm Service.

MCS – vacuum too low.

Check connectors. If they are ok:

Contact the Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 4.15.4, page 97)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 64)
.