Physical and environmental specifications, Complementary cumulative – Bird Technologies 5019D-Manual User Manual
Page 51

39
Complementary Cumulative Distribution Function
(CCDF)
Measurement Range*
0.1 – 100%
Measurement Uncert.
± 2%
Threshold Level Range 25 mW to 60 W
Level Set Accuracy
As peak power uncert. + 2%
* Duty Cycle and CCDF read out dependent on display
method.
Physical and Environmental Specifications
Temp, Operating
–10 to +50 °C (+14 to +122 °F)
Temp, Storage
–40 to +80 °C (–40 to +176 °F)
Mechanical Shock
and Vibration
MIL-PRF-28800F class 3
Humidity, Max
95% (non-condensing)
Altitude, Max
15,000 ft. (4,500 m)
Dimensions, Nominal
4.75” x 4.6” x 1.3” (121 x 117 x 33 mm)
Weight, Max
1.2 lb. (0.55 kg)
EMC
EMC Directive (2004/108/EC)
European Standard: EN 61326—
Electrical Equipment for
measurement, control and laboratory
use; EMC Requirements
Test Spec (for radiated immunity): EN
61000-4-3—Testing and
measurement techniques - 10V/meter