5000-xt – Bird Technologies 5019D-Manual User Manual
Page 18

6
If calibration fails, “Cal Fail” will be displayed.
a.
Press a key to return to normal operation,
b.
Ensure that the WPS is properly connected, and the
RF is off.
c.
Rezero. Go back to Step 3.
5000-XT
1.
Make sure the sensor has reached a stable operating
temperature.
2.
Make sure no RF power is applied to the sensor.
3.
Press and hold “0” for two seconds to begin zero cali-
bration.
WARNING
Do not interrupt the calibration.
Note:
“Zero/Cal” will be displayed and calibration
will begin.
Note:
Zeroing the sensor takes 60 seconds. The
bar graph will display calibration progress.
Note:
When complete, “PASS” should be displayed.
4.
Press Enter to return to normal operation when zeroing is
complete.
Note:
If “FAIL” is displayed, make sure no RF power
is applied to the sensor and perform the procedure
again.