Chapter 5

Maintenance

This chapter describes routine maintenance, along with troubleshooting instructions for the power meter and 
power sensor. Disassembly instructions for the Bird Cal Cart are also provided. For service beyond this level, return 
the unit to a qualified service center

The Bird 4421 Power Meter requires only simple, routine maintenance.

Wipe off dust and dirt regularly.

Check the connectors and cables for damage.

Clean the connector contacts with alcohol or dry cleaning solvent.

Troubleshooting

Since the power meter and power sensor can only work together, the first step is to determine which is malfunction‐
ing. Connect the power sensor to the meter and perform the functional test on page 13. If the power meter is mal‐
functioning, refer to the troubleshooting table below. If the power sensor is malfunctioning, return it for service.

This manual cannot list all malfunctions that may occur, or corrective actions. If a malfunction is not listed or is not 
corrected by the listed corrective actions, contact a qualified service center.

Note:

For all power sensor tests, following the test procedure will check the sensor not covered by the 

safety cover. To test the other sensor slide the safety cover over, then proceed as normal.

WARNING

To avoid personal injury, disconnect the power cord from the AC line before performing any maintenance, 

including fuse replacement.

WARNING

Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being 

applied. Leaking RF energy is a potential health hazard.

WARNING

The Bird 4421 contains no user‐serviceable parts. Do not remove its cover.

CAUTION

Due to the complexity of the Bird Power Sensor, field repairs beyond general maintenance should not be 

attempted. Removal or disturbance of the power sensor cover can result in cancellation of lifetime warranty.