7

Chapter 3

Operating Instructions

Normal Operation

Note:

After setting up the Bird Cal Cart:

1.

Connect the unit to the AC line, if necessary.

2.

Press the ON/OFF button on the Bird 4421. 

Note:

The power up display will flash “4421” for approximately 10 seconds 

and then change to the normal operating display.

Operating Loads with Attached Blowers

Note:

For the 8931 and other loads with attached blowers ONLY.

1.

Check that the indicator light is on.

2.

Check that the switch is set to MANUAL.

3.

Apply RF power.

4.

Make measurements. 

Note:

See "4421 Operating Instructions" on page 9 for specific instructions on 

controlling the meter.

Shutdown

1.

Turn off RF power at the source.

2.

Press the ON/OFF button on the 4421.

Note:

For blower‐equipped loads, wait approximately 15 minutes or for the 

fans to stop running. This will allow the load to cool without causing heat stress.

3.

Disconnect the AC line.

WARNING

Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being 

applied. Leaking RF energy is a potential health hazard.

WARNING

Refer to the supplied load manual for load‐specific warnings and cautions.

CAUTION

Maximum power dissipation is severely reduced when the blower is not running. If the 

indicator light should turn off, immediately reduce RF power by 75%.