Specifications – Yamato Scientific Vacuum controller Vacuum controller User Manual
Page 30
28
Specifications
Model
VR100
Pressure Control Range(hPa):
Reached pressure in the vacuum system ~ 800
±5% Full scale
Pressure Measurement/Display
Range(hPa):
1~1013
±5% Full scale
Pressure Setting Range(hPa)
1~1013
Grading Pressure
Setting Range(hPa):
66~600
Grading Time Setting Range
1~99hrs.59min.
Measurement Calibration
Setting Range(hPa):
Measurement
±5% Full scale
Pressure Control Method:
Feedback ON/OFF Control by Semiconductor Pressure Sensor
and Control Valve
Pressure Setting Method:
Digital Setting by
▲▼(up/down) Key
Pressure (etc.) Indication
Method:
Measured Pressure, Set point (pressure, ramp time) are digitally
displayed on the indicator
Setting/Displayed Pressure
Resolving Power (hPa):
1
Functions For Operation:
Fixed Control Operation(Step-Control ranged1~800hPa), Ramp
Control Operation (ramp control ranged66~600hPa), Free Mode
Operation ( cancels operations. Depends on reached pressure
in the vacuum system)
Control Mode:
Fixed Value Control, Ramp Control
Auxiliary Functions:
Measured Pressure Calibrating Function(fixed value
control; accepted before or during the free mode), Ramp Control
Set point Edit Function
Controller:
IV LE type Controller +PIO 4 Indicator
Pressure Sensor:
Measurement Range;0~1030hPa
±5% Full scale
Solenoid Valve for Control:
AC100V normally close
Leak Valve:
AC100V normally open
Safety System:
Fuse (rating;AC250V 2A), Leak Function, Self-Diagnostics
Sensor (for disconnectioned sensor)
Dimensions (WxDxH mm):
155x190x145(approximately)
Power Supply:
AC100V 1A
Weight:
3.5kg (approximately)
Accessories:
Metal Fixture, Screws, Fuse(1), Guarantee, Instruction Manual
Optional Accessories:
None